Comparison of Cut Out Transmission Variations on a 10x10 cm2 Applicator for Linac Electron Beam

Alisya Oktafianingrum, Robert Janssen Stevenly, Pratiwi Sri Wardani, Erlinda Ratnasari Putri

Abstract


Radiotherapy has provisions for providing radiation doses for patient therapy, as giving the minimum possible dose to healthy tissue and giving the maximum possible dose to cancer. The aim of this research is to compare the electron beam output with measurements of cut out transmission variations (using a built-in block, without using a block, and using a cerrobend block). This research used a 10x10 cm2 applicator which was attached to the Linac. The blocks were installed and the absorbed dose was measured using water phantom and Markus ionization detektor. Besides that, absorbed dose was also measured without blok. The energy used in this study was 12 MeV and the a mount of radiation dose from Linac was 100 MU. Measurement results using a built-in block (lead) of 100 cGy, a cerrobend block of 100 cGy. Measurements without using a block were 102 cGy. This measurement has a comparison in the results that get low and high deviation values. Measurements using lead blocks and cerrobends are 0%, while measurements without blocks have a deviation value of 2%. These three measurements are still within the IAEA 398 TRS standard tolerance limit, i.e, ± 2%.


Keywords


Applicators; Blocks; Electrons; Radiotherapy; TRS 398

References


Alvia, H. A. (2022). Pengukuran Dosis Radiasi pada Phantom di CT-SCAN. Karya Tulis Ilmiah. Universitas Widiya Husada Semarang.

Badan Pengawas Tenaga Nuklir. (2013). Peraturan Kepala Badan Pengawas Tenaga Nuklir Nomor 3 Tahun 2013 tentang Ketentuan Keselamatan Desain Reaktor Daya. Perka BAPETEN, 3 Thn 2013, 1–41.

Fitriatuzzakiyyah, N., Sinuraya, R. K., & Puspitasari, I. M. (2017). Cancer Therapy with Radiation: The Basic Concept of Radiotherapy and Its Development in Indonesia. Indonesian Journal of Clinical Pharmacy, 6(4), 311-320.

Heru, N. S., Saraswati, A., Tarigan A., & Samsun (2017). Verifikasi Nilai Dosis terhadap Variasi Lapangan Blok Elektron pada Energi 6 MeV dan 9 MeV Pesawat LINAC Varian (CLINAC IX-SN4740), Prosiding Pertemuan dan Persentasi Ilmiah Penelitian Dasar Ilmu Pengetahuan dan Teknologi Nuklir, Editor: Kusminarto, 293-298, Yogyakarta.

IAEA. (2000). IAEA TRS 398:Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water. International Atomic Energy Agency, Technical Report Series, TRS-398, Vienna : IAEA, 183.

Khan, F. M. (2003). Physics of Radiation Therapy Third Edition. The Journal of the American Medical Association, 1138.

Mather, S. J., & Mansi, L. (2008). IAEA Technical Report Series. Eurpean Journal of Nuclear Medicine and Moleculer Imaging, 35(5), 1030-1031.

Mayles, P., Nahum, A., & Rosenwald, J. C. (2007). Handbook of radiotherapy physics: Theory and practice. In Handbook of Radiotherapy Physics: Theory and Practice (pp. 1–1453).

Podgorsak E.B. (2005). Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency, Technical Editor, Vienna. Thirth Edition.

Purba, S. (2010). Dosis Transmisi Berkas Sinar-X 6 MV untuk Lapangan Tidak Teratur dengan Variasi Blok. Tesis. Deparmen Fisika Universitas Indonesia, Depok.

Puspitasari, R. A. (2020). Analisis Kualitas Berkas Radiasi LINAC Untuk Effektivitas Radioterapi. Jurnal Biosains Pascasarjana, 22(1), 11-19.

Ramona, F., Milvita, D., & Diyona, F. (2020). Verifikasi Nilai Dosis Radiasi Berkas Elektron Pesawat LINAC dengan Luas Lapangan Blok Cerrobend (6x6) cm2 dan (8x8) cm2 Berdasarkan Protokol IAEA TRS 398. Jurnal Fisika Unand, 9(2), 238-243.

Santoso, H., Hamdi, M., & Saktioto, S. (2020). Analisis Kualitas Data Keluaran Harian Elektron Akselerator Linear. Komunikasi Fisika Indonesia, 17(3), 166-171.

Septiyanti, I., Khalif, M. A., & Anwar, E. D. (2020). Analisis Dosis Paparan Radiasi Pada General X-Ray II Di Instalasi Radiologi Rumah Sakit Muhammadiyah Semarang. Jurnal Imejing Diagnostik (JImeD), 6(2), 96-102.

Setiawati, E., Ariyanty E., Ramantisan, S., & Hidatullah, M. (2018). Analysis of Size and Location of Irradiation Block of Electron Linear Accelerator to Radiation Output and Absorbed Dose, International Journal of Scientific Engineering and Science 2 (11), 35-38

Suharmono, B. H., Anggraini, I. Y., Hilamaniyya, H., & Astuti, S. D. (2020). Quality Asurance (QA) dan Quality Control (QC) pada Pesawat Linac. Jurnal Biosains Pascasarjana, 22(2), 73-80

Vadila, M., & Milvita, D. (2018). Analisis Keluaran Berkas Elektron Pesawat Terapi LINAC Tipe Varian CX 6264 di Rumah Sakit Universitas Andalas. Jurnal Fisika Unand, 7(2), 91-96.




DOI: http://dx.doi.org/10.20527/flux.v21i2.17701

Article Metrics

Abstract view : 584 times
PDF (Bahasa Indonesia) - 554 times

Refbacks

  • There are currently no refbacks.


Copyright (c) 2024 Jurnal Fisika Flux: Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat

Creative Commons License
This work is licensed under a Creative Commons Attribution-NoDerivatives 4.0 International License.

Association with:

Physical Society of Indonesia

Indexed by:

 

Creative Commons License
Jurnal Fisika FLux: Jurnal Ilmiah FMIPA Universitas Lambung Mangkurat is licensed under a Creative Commons Attribution-NoDerivatives 4.0 International License.